STRAIN GAUGE (BRIDGE) INPUT TRANSMITTERS
PLUG-IN MODULE, FIXED RANGE

JH4051I: Strain Gauge Input, Input/Output Isolated
- Input Spans as low as 0.4mV/V (at 10V excitation)
- Excitation Adjustable from 4 to 12V
- Quick-Check Red/Green Output LEDs
- Uses Standard 11-Pin Socket
- Nonstandard Ranges Available
- AC or DC Power Options
|
JH4051I
PDF
DATA SHEET |
Model JH4051I supplies DC excitation to a strain gauge or other resistance
bridge, amplifies the bridge output and provides a proportional DC voltage or current.
A fixed-range device, it is precisely calibrated to your specified range at the factory.
Red-green Quick-Check LEDs give a quick indication of the relative output. Red is brighter
at the low end, green at high, while at mid-scale both are approximately equal. Red-only indicates offscale
low while green-only indicates offscale high.
Available options include AC or DC power choices
and reverse-action Option RT (decreasing output with increasing input).
ORDERING INFORMATION
|
Model Number: |
Select JH4051I-AC for AC power, JH4051I-DC for DC power.
|
|
Bridge Excitation: |
Specify any voltage between 4 and 12Vdc. Will
be factory set to 10V if not specified on your order. User-adjustable. (Note: 40mA maximum load. For 120ohm
bridges, specify 4.8V or less.)
|
|
Input Range: |
Specify any input range allowed by
the "Input Capabilities" spec, below. Specify either as millivolts per volt, or as millivolts. (Millivolts
equals millivolts-per-volt multiplied by the excitation voltage. For example, 0-2.0 mV/V at 10V excitation
equals 0-20mV.) |
| Output Range: |
Specify any DC voltage or current range allowed by
the "Output Capabilities" spec, below.
|
|
Power: |
Specify 115Vac, 230Vac, 12Vdc or 24Vdc.
|
|
Reverse-Acting Transmitter (decreasing output with increasing input): |
Specify Option RT.
|
|
Loop-Powered Output: |
4/20mA "current sink" output stage
for connection to devices whose inputs provide 24Vdc loop excitation. Specify Option LPO.
|
INSTALLATION AND CONNECTIONS
The JH4051I plugs into any standard 11-pin circular ("octal") relay socket.
JH Technology offers two sockets: DS011 for DIN-rail or surface mounting and ST011 for Snap-Track mounting
(see the Accessories page). Pin connections are:
Pin 1: Power (AC or, if DC power option, DC plus).
Pin 2:
No connection.
Pin 3: Power (AC or, if DC power option, DC minus).
Pin 4: Input plus (bridge
output).
Pin 5: Input minus (bridge output).
Pin 6: No Connection
Pin 7: Bridge excitation plus.
Pin 8: Bridge
excitation minus.
Pin 9: Output plus.
Pin 10: Output minus.
Pin 11: No connection.
DETAILED SPECIFICATIONS
|
Input Capabilities: |
4mV minimum span (0.4mV/V at 10V excitation).
10V maximum span (1V/V at 10V excitation). Offset ranges are allowed.
(Input Impedance: 200kohms.)
|
|
Excitation Voltage: |
Adjustable from 4 to 12Vdc. 40mA maximum
load. Factory set to 10Vdc unless otherwise specified. (Note: 120 ohm bridges limited to 4.8V or less. 350
ohm bridges are ok to 12V.)
|
|
Voltage Output Capabilities: |
1 volt minimum output span, -10 to
+15V absolute limit. Offset ranges are allowed. Maximum output load, 10mA (1Kohm at 10V output).
|
|
Current Output Capabilities: |
1mA minimum output span, 0 to +25mA
absolute limit. Positive offsets are allowed, negative outputs are not. Output drive capability, 24V (1,200
ohms max. at 20mA output).
|
|
Accuracy (factory calibration): |
+/-0.1% of span or 10 microvolts, whichever is greater.
|
|
Adjustability: |
Zero and span each are adjustable approx. +/-15%
of span.
|
|
Linearity: |
+/-0.05% of span or better.
|
|
Response Time: |
Under 100 milliseconds.
|
|
Isolation: |
Power, 1,500Vac rms (2,100V peak). Input/Output, 1,000Vac rms (1,400V peak).
|
|
Guaranteed Operating Temperature: |
-10 to +60 deg. C (14 to 140
deg. F).
|
|
Temperature Stability: |
+/-(0.02% of span plus 1.3 microvolts) per
deg. C, or better.
|
|
Power Requirements: |
AC, 115 or 230Vrms, 50/60Hz, 2.5V-A. DC, 12
or 24V, 2.5W.
|
|