JH5200 THERMOCOUPLE INPUT TRANSMITTER
DIN-RAIL MOUNT, FIXED RANGE

- Available for All Thermocouple Types
- Upscale Burnout Standard, Downscale Optional
- Automatic Cold-Junction Compensation
- Quick-Check Red/Green Output LEDs.
- DIN-Rail Mount, 1 Inch Wide
- AC or DC Power Options
|
JH5200
PDF
DATA SHEET |
The Model JH5200 Thermocouple Input Transmitter provides an isolated DC
output proportional to a thermocouple input. Output is linear with millivolt
input. Automatic cold junction compensation and a low-drift input amplifier
maintain accuracy under varying ambient conditions.
Red-green Quick-Check LEDs give a quick indication of the relative output.
Red is brighter at the low end, green at high, while at mid-scale both are
approximately equal. Red-only indicates offscale low while green-only indicates
offscale high.
Available options include AC and DC power choices, reverse-acting transmitter
(decreasing output with increasing input) and downscale burnout indication
(upscale is standard).
ORDERING INFORMATION
| Model Number: |
Specify Model JH5200A for AC power,
JH5200D for DC power. |
| Thermocouple: |
Specify any standard thermocouple
type (E, J, K, T, R, S, B or N), or contact factory for others. |
| Input Range: |
Specify range required in deg. C or
deg. F. (See Specifications for input capabilities.) |
| Burnout Indication: |
Upscale burnout indication is
standard. (Output goes high if thermocouple input opens.) If downscale
indication is required, specify "Downscale Burnout" on your
order. |
| Output Range: |
Specify any DC voltage or current
range allowed by the "Output Capabilities" spec, below. |
| Power: |
Specify 115Vac, 230Vac, 12Vdc or
24Vdc. |
| Reverse-Acting Transmitter (decreasing
output with increasing input): |
Change last digit of the model number to 1
(for example, JH5201A). |
| Loop-Powered Output: |
4/20mA "current sink" output stage
for connection to devices whose inputs provide 24Vdc loop excitation.
Change the last digit of the model number to 2 (for example, JH5202A). |
| Urethane Coating: |
Specify Option U. |
Model JH5200 snaps onto 35mm DIN rail. Connections are made to the
front-panel terminals. The terminal strip unplugs to facilitate calibrating or
replacing the transmitter.
Connections to the 8 terminals (top to bottom) are:
1: Thermocouple plus.
2: Thermocouple minus.
(Note: On most thermocouples, red is
minus.)
3: No connection.
4: No connection.
5: Output plus.
6: Output minus.
7: Power (AC or, if DC power option, DC plus).
8: Power (AC or, if DC power option, DC minus).
DETAILED SPECIFICATIONS
| Input Thermocouple: |
Available for types E, J,
K, T, R, S, B or N. Contact factory for others. |
| Input Capabilities: |
Any span 4mV or higher.
(4mV equals 100 deg. C for type K thermocouple.) Offset ranges are
allowed. |
| Voltage Output Capabilities: |
1 volt minimum output
span, -10 to +15V absolute limit. Offset ranges are allowed. Maximum
output load, 10mA (1Kohm at 10V output). |
| Current Output Capabilities: |
1mA minimum output span,
0 to +25mA absolute limit. Positive offsets are allowed, negative
outputs are not. Output drive capability, 24V (1,200 ohms max. at 20mA
output). |
| Accuracy (end-point,
factory calibration): |
+/-0.1% of span. On low
input spans, +/-0.25 deg. C or 20 microvolts, whichever is greater. |
| Adjustability: |
Zero and span each are
adjustable approx. +/-15% of span. |
| Response Time: |
Under 100 milliseconds. |
| Isolation: |
3-way
(Power/Input/Output) 1,500Vac rms (2,100V peak). |
| Guaranteed Operating
Temperature: |
-10 to +60 deg. C (14 to
140 deg. F). |
| Temperature Stability: |
+/-(0.02% of span plus
1.3 microvolts) per deg. C, or better. |
| Power Requirements: |
AC, 115 or 230Vrms,
50/60Hz, 2.5V-A. DC, 12 or 24V, 2.5W. |
|